Current Sensors
Digital Horizontal/Vertical Interface Board (DHV) OEM
Digital Horizontal/Vertical Interface Board (DHV) is used on combination vertical/horizontal sashes for OEM partner.
Please sign in to access more documents
Once signed in, you may be able to access additional documents for your account.
Phoenix Controls Sash Sensors detect a fume hood’s sash position. Sensors can be configured to operate with the sash configurations found on most VAV fume hoods. The sash sensors are used together with a Phoenix Controls fume hood monitor and airflow control valve to maintain a constant average face velocity at the sash opening
Features & Benefits:Material: compatible with PPC flame retardant black box part number 520-000-011LF- Enclosure:
- Dimensions: 1.9"" H x 6"" W x 3.2"" D (48 x 12 x 81 mm)
- Weight with PCB: > 6 oz. (170 grams)
- Color: black
- Power requirements with 24 VAC input = ± 15%; 50/60 Hz; 6 VA, or ± 15 VDC input = ± 15%; 200 mA
- 8 Inputs
- Voltage: 0.0
- 10.5 Vdc; ±1% FS; jumper position = OUT, or
- Resistance: 0 - 10K ohms; ±1% FS;
- jumper position = IN (default)
- 2 Outputs
- OUT: 0.0 - 10.0 Vdc; 1% FS (Total Sash Signal)
- EMERG: Unused
- Operating Range
- Ambient Temperature: 0 to 50 C (32 to 122 F)
- Storage Temperature: -20 to 70 C (-4 to 158 F)